Application overview

The Guidebook, GB 5085 Identification Standard for Hazardous Waste-Identification for Toxic Substance Content, stipulates the limit values for over 20 toxic and hazardous elements. However, the coverage of diversified types of wastes and hazardous elements, the complexity of the sample matrix and the wide range of contents have posed challenges to the instrument analysis. Moreover, it is difficult for the traditional XRF relying on the standard samples to carry out the quantitative analysis.

High-sensitivity XRF (HS XRF) with the Fast Fundamental Parameter method (Fast FP) is well responsive to the heavy metal detection in solid and hazardous waste, with many distinct advantages such as less standard samples, easier sample processing, and faster testing, which make it highly adaptable to all kinds of waste samples.




Method principle


 Invention patent

Core technology on Hardware:High-sensitivity XRF(HS XRF®

1) Monochromatic excitation
The background interference caused by the X-ray tube emission spectrum after being scattered by the sample can be obviously decreased.

2) Focused excitation
The Energy can be focused, which distinctly increase the intensity of fluorescence radiation of the sample element received by the SDD detector.


Core technology on Software:Fast Fundamental Parameter method(Fast FP®






 Copyrights of the software:

1) A mathematical model and database were established for the calculation of the X-ray process.

2) The matrix, spectral interference, and background subtraction can be calculated correctly.

3) Accurate quantitative results can be obtained with a few of the standards.











                                                              ​Fig.2. Fitting of solid-waste samples spectrums                                           


Tab.1.  Test results of solid-waste samples

Elements
Certified value
Found value
Relative deviation
(Sb)
88.9±8.1
79.37
10.71%
(As)
40.2±4.1
47.01
16.93%
(Ba)
221±20
213.53
3.38%
(Cd)
12.5±1.3
12.50
0.01%
(Cr)
244±21
221.56
9.20%
(Pb)
78.3±7.3
79.33
1.32%
(Co)
56.1±5.2
53.20
5.17%
(Cu)
200±18
190.36
4.82%
(Mn)
409±38
427.12
4.43%
(Ni)
83.8±7.9
88.33
5.40%
(Sr)
298±25
296.40
0.54%
(Sn)
118±10
116.78
1.03%
(Hg)
118±10
99.28
15.86%
(Zn)
164±15
143.49
12.51%
(Mo)
128±11
126.68 1.03%



Method Detection limits
Detection of limits for Hazardous Elements in Solid Waste
                                                                                                                        unit:mg/kg

Types of Elements Light Elements Light Metals Metals Heavy Metals Ⅰ Heavy Metal Ⅱ
Elements S、Cl、P  K、Ca、V Cr、Mn、Co Hg、Ag、Ba、Sr、
Pt、Pa
 Ni、Cu、Zn、As、
Pb、Br、Se、Cd、
Sb、Tl
Limit of Detection 50 10 5 2 0.2


Comparison of Heavy Metal Limits and Limits of Detection stipulated in the GB 5085 Identification Standards for Hazardous Wastes-Identification for Substance Content
Heavy Metals Cu\Zn\Ba Cd\Se Pb\Cr\Ni\Ag\As Hg
Limits(mg/L) 100 1 5 0.1
Limits of Detection (mg/L) 0.5 0.1 0.2 0.3
 





Characteristics of the Method

 Broad elements
   More than 20 toxic elements specified in the identification standards of hazardous waste can be analyzed at one time, and the detection limit meets the criteria of determination;

 Wide adaptation
   Matrixes of different kinds of solid waste can be accurately calibrated by the Fast FP, which is suitable for various solid, powder and liquid samples;

 Accuracy
   Matrix differences and spectrums interference can be eliminated by the Fast FP, and we can use a few of standard samples to correct the calculation error;

 Rapidity
   Sample processing is simple and fast, and the analysis time is less than ten minutes per sample;

 On-site Analysis
   PHECDA-ECO is light and easy to carry, which can easily complete the on-site measurement of toxic elements in solid waste and hazardous waste

 Sample processing
    We provide the complete sample processing equipment and operating procedures;






Declaration:
Ancoren is the owner of the above patents for technological inventions. We reserve the right to defend rights in case of any infringement.

For more details, please contact the staffs!